Catalog Number C45091342-NLJ
Device Problem Leak/Splash (1354)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/12/2019
Event Type  malfunction  
Event Description
Information was received indicating that an air leak alarm occurred following connecting a smiths medical portex® disposable anesthesia breathing circuit to an anesthesia device.There were no reported adverse patient effects.
 
Manufacturer Narrative
One sample was returned for evaluation.Visual inspection of the device found it to be in good physical condition.The circuit underwent functional testing; device was noted to be leaking.The customer complaint has been confirmed, and the problem source has been determined to be manufacturing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX® DISPOSABLE ANESTHESIA BREATHING CIRCUIT
Type of Device
ANESTHESIA BREATHING CIRCUIT KIT (ADULT & PEDIATRIC)
Manufacturer (Section D)
SMITHS MEDICAL ASD; INC.
6000 nathan lane n
minneapolis, MN 55442
MDR Report Key9613564
MDR Text Key180317496
Report Number3012307300-2020-00296
Device Sequence Number1
Product Code OFP
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 03/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue NumberC45091342-NLJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/25/2019
Initial Date Manufacturer Received 12/23/2019
Initial Date FDA Received01/21/2020
Supplement Dates Manufacturer Received02/18/2020
Supplement Dates FDA Received03/19/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-