MAUDE Adverse Event Report: NEUROPACE, INC. NEUROPACE RNS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEUROPACE, INC. NEUROPACE RNS SYSTEM Back to Search Results
Model Number DL-330-10
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Impaired Healing (2378); Patient Problem/Medical Problem (2688); No Code Available (3191)
Event Date 12/16/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
In (b)(6) 2019, the patient noted a protrusion and some fluid leaking around a portion of the rns system incision site.The patient was scheduled for a revision procedure on (b)(6) 2020.During the procedure, it was determined that one of the unconnected, capped leads was applying outward pressure near the incision line, preventing proper healing.The neurosurgeon re-positioned the lead, securing it away from the incision site.No further information was provided by the treating center.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUROPACE RNS SYSTEM
Type of Device
NEUROPACE RNS SYSTEM
Manufacturer (Section D)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer (Section G)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer Contact
ramona gonis
455 n. bernardo ave.
mountain view, CA 94043
6502382788
MDR Report Key9614122
MDR Text Key187169924
Report Number3004426659-2020-00002
Device Sequence Number1
Product Code PFN
UDI-Device Identifier00855547005045
UDI-Public010085554700504517220411
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 01/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator No Information
Device Model NumberDL-330-10
Device Catalogue Number1007605
Device Lot Number27875-1-1-1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/03/2020
Initial Date FDA Received01/21/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age31 YR
-
-