• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ALLURE QUADRA RF CRT-P; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ALLURE QUADRA RF CRT-P; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM3242
Device Problems Pacemaker Found in Back-Up Mode (1440); Wireless Communication Problem (3283)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/06/2020
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that the patient attempted to send remote transmission of the pacemaker data to the clinic.However, the pacemaker failed to communicate the transmitter using radio frequency (rf) telemetry.The patient was sent a new transmitter with inductive telemetry on (b)(6) 2020.While attempting to pair the transmitter, the pacemaker went into backup vvi operation.On (b)(6) 2020, normal pacemaker operation was successfully restored and the transmitter was able to retrieve and send pacemaker data to the clinic.There were no adverse consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURE QUADRA RF CRT-P
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
MDR Report Key9618213
MDR Text Key176377527
Report Number2017865-2020-00836
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734507912
UDI-Public05414734507912
Combination Product (y/n)N
PMA/PMN Number
P030035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial
Report Date 01/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/22/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date06/30/2017
Device Model NumberPM3242
Device Catalogue NumberPM3242
Device Lot NumberA000015930
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received01/06/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/16/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age76 YR
-
-