• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE Back to Search Results
Model Number EL5ML
Device Problem Misfire (2532)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/28/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # t94j0l.Investigation summary: the analysis results found that the el5ml device was returned with no damage in the external components.In addition, 2 malformed clips were returned inside of a plastic bag.In an attempt to replicate the reported incident, the instrument was tested for functionality.During the analysis, the device was cycled and it fed and formed 1 conforming clip.The event described could not be confirmed as the device performed without any difficulties noted.The reported complaint could not be confirmed.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformance's related to the reported complaint condition were identified.
 
Event Description
It was reported that during an unknown procedure, the clip applier misfired within the patient.Clip applier was removed without event.Surgery was delayed (number of minutes delayed was unknown) and another device was obtained for successful completion of the procedure.There was no reported patient consequence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX-5MM ENDO CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kara ditty-bovard
475 calle c
guaynabo 00969
*   00969
6107428552
MDR Report Key9618411
MDR Text Key204397542
Report Number3005075853-2020-00513
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036001843
UDI-Public10705036001843
Combination Product (y/n)N
PMA/PMN Number
K050344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/22/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEL5ML
Device Catalogue NumberEL5ML
Device Lot NumberT94J0L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/21/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/21/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/14/2019
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-