(b)(4).Batch # t94j0l.Investigation summary: the analysis results found that the el5ml device was returned with no damage in the external components.In addition, 2 malformed clips were returned inside of a plastic bag.In an attempt to replicate the reported incident, the instrument was tested for functionality.During the analysis, the device was cycled and it fed and formed 1 conforming clip.The event described could not be confirmed as the device performed without any difficulties noted.The reported complaint could not be confirmed.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformance's related to the reported complaint condition were identified.
|