MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 RECLAIM ASSEMB TOOL ADAPT 85MM; RECLAIM INSTRUMENTS : INSERTION DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 RECLAIM ASSEMB TOOL ADAPT 85MM; RECLAIM INSTRUMENTS : INSERTION DEVICES Back to Search Results
Model Number 2975-00-610
Device Problems Component Missing (2306); Material Twisted/Bent (2981)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/01/2020
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the 85 mm assembly rod adaptor is missing locking bearings.75mm, 85mm, and cap do not screw together properly, need new products found defective during training in office.
 
Event Description
Additional information received indicating that it appears the locking mechanism inside the adaptor was stripped, had no bearings.
 
Manufacturer Narrative
Product complaint # : (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary examination of the returned instrument confirmed the complaint.Depuy-synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot null device history batch null device history review null if information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RECLAIM ASSEMB TOOL ADAPT 85MM
Type of Device
RECLAIM INSTRUMENTS : INSERTION DEVICES
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key9619784
MDR Text Key189349972
Report Number1818910-2020-02577
Device Sequence Number1
Product Code LXH
UDI-Device Identifier10603295155966
UDI-Public10603295155966
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 01/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2975-00-610
Device Catalogue Number297500610
Device Lot NumberSO2002961
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/06/2020
Initial Date Manufacturer Received 01/10/2020
Initial Date FDA Received01/23/2020
Supplement Dates Manufacturer Received02/06/2020
03/05/2020
Supplement Dates FDA Received02/27/2020
03/06/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-