• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. HARMONIC SCALPEL HAND PIECE; INSTRUMENT, ULTRASONIC SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. HARMONIC SCALPEL HAND PIECE; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Catalog Number HP054
Device Problems Device Displays Incorrect Message (2591); Output Problem (3005)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/29/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # r94u5f.Investigation summary: the device was received with no apparent damage.It was connected to a generator, evaluated with a test instrument and was found to be functional.The instrument was disassembled to inspect the internal components.The moisture indicator was positive.The handpiece has a number of seals to prevent fluids from entering the housing."positive moisture indicator¿ describes a condition where water enters the handpiece cavity during the steam sterilization process.The primary path of ingress is the distal seal, this may be caused by a reduction of the compressive force on the distal seal.However no definitive root cause could be drawn.It is probable that the ingress of moisture affected handpiece functionality.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this batch.
 
Event Description
It was reported that during an unknown procedure, the harh45 did not work.They verified the hand piece and it still had uses available.When they change the energy device to another harh45, that one worked fine, but the first one gave an error on the gen11.No patient consequence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HARMONIC SCALPEL HAND PIECE
Type of Device
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kara ditty-bovard
475 calle c
guaynabo 00969
*   00969
6107428552
MDR Report Key9626754
MDR Text Key204400449
Report Number3005075853-2020-00526
Device Sequence Number1
Product Code LFL
UDI-Device Identifier10705036002147
UDI-Public10705036002147
Combination Product (y/n)N
PMA/PMN Number
K002906
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 01/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/24/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberHP054
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/09/2020
Date Manufacturer Received01/10/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/20/2018
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-