Model Number 595000-001
Device Problem Audible Prompt/Feedback Problem (4020)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/30/2019
Event Type  malfunction  
Manufacturer Narrative
The freedom driver has been returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The customer, a syncardia certified hospital, reported that the freedom driver exhibited a fault alarm while supporting a patient.The patient was switched to a backup driver and there was no reported adverse patient impact.
 
Manufacturer Narrative
The driver's alarm history was reviewed and revealed one new fault code.A '4b' fault code alarm is typically produced when the driver is disconnected from the patient and most likely occurred when the driver was being switched off the patient.Only permanent alarms are recorded in the history; intermittent and/or recoverable alarms would not be recorded.The driver passed all sections of functional testing.Additionally, an onboard battery discharge/ recharge test and an onboard battery exchange test were performed.The driver functioned as intended with no alarms produced during both tests.During investigation testing, the customer-reported issue was not able to be reproduced and there was no evidence of a device malfunction.It is possible the customer-reported alarm was a battery alarm.A battery alarm is recoverable and would not be recorded in the driver's history unless the alarm was unaddressed for 4 minutes and 15 seconds.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.Ce 5136 follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key9627248
MDR Text Key177378325
Report Number3003761017-2020-00030
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003121
UDI-Public(01)00858000003121
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 02/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number595000-001
Device Catalogue Number595000-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/06/2020
Initial Date Manufacturer Received 12/30/2019
Initial Date FDA Received01/24/2020
Supplement Dates Manufacturer Received12/30/2019
Supplement Dates FDA Received02/06/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age48 YR
-
-