• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION 2 DRIVER; EXTERNAL PNEUMATIC DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION 2 DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Model Number 397002-001
Device Problem Audible Prompt/Feedback Problem (4020)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/20/2019
Event Type  malfunction  
Manufacturer Narrative
The companion 2 driver has been returned to syncardia.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The customer, a syncardia authorized distributor, reported that the companion 2 driver exhibited an increase of the pneumatic pressure leading to multiple right over pressure alarms while supporting a patient.The patient was switched to a backup driver and there was no reported adverse patient impact.
 
Manufacturer Narrative
The customer-reported issue was confirmed and reproduced during investigation testing.The root cause was determined to be a malfunction of both the left and right electronic pressure regulators as they gave an output voltage outside the required range.Syncardia has a corrective and preventive action (capa) to address incorrect pressure alarms related to the electronic pressure regulator.Syncardia has completed its evaluation of this complaint and is closing this file.Ce 5139 follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA COMPANION 2 DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson, az AZ 85713
MDR Report Key9627251
MDR Text Key177378639
Report Number3003761017-2020-00031
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003107
UDI-Public(01)00858000003107
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Type of Report Initial,Followup
Report Date 02/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/24/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number397002-001
Device Catalogue Number397002-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/09/2020
Date Manufacturer Received12/31/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age34 YR
-
-