• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA; EXTERNAL PNEUMATIC DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Model Number 397002-001
Device Problem Audible Prompt/Feedback Problem (4020)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/29/2019
Event Type  malfunction  
Manufacturer Narrative
The companion 2 driver has been returned to syncardia.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The customer, a syncardia certified hospital, reported that the companion 2 driver exhibited an unrecoverable high temperature alarm while supporting a patient.The patient was switched to a backup driver and there was no reported adverse patient impact.
 
Manufacturer Narrative
During investigation testing, the driver passed all sections of functional testing without any alarms or abnormalities.The customer-reported temperature alarm was not reproduced and the driver performed as intended.The root cause of the customer-reported issue could not be conclusively determined and there was no evidence of a device malfunction.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.(b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson, az AZ 85713
MDR Report Key9627273
MDR Text Key177378553
Report Number3003761017-2020-00025
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003107
UDI-Public(01)00858000003107
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 06/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number397002-001
Device Catalogue Number397002-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/09/2020
Initial Date Manufacturer Received 12/29/2019
Initial Date FDA Received01/24/2020
Supplement Dates Manufacturer Received12/29/2019
Supplement Dates FDA Received06/22/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age57 YR
-
-