Model Number 397002-001 |
Device Problem
Audible Prompt/Feedback Problem (4020)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/29/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
The companion 2 driver has been returned to syncardia.The results of the investigation will be provided in a follow-up mdr.(b)(4).
|
|
Event Description
|
The customer, a syncardia certified hospital, reported that the companion 2 driver exhibited an unrecoverable high temperature alarm while supporting a patient.The patient was switched to a backup driver and there was no reported adverse patient impact.
|
|
Manufacturer Narrative
|
During investigation testing, the driver passed all sections of functional testing without any alarms or abnormalities.The customer-reported temperature alarm was not reproduced and the driver performed as intended.The root cause of the customer-reported issue could not be conclusively determined and there was no evidence of a device malfunction.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.(b)(4) follow-up report 1.
|
|