Model Number 2426-0500 |
Device Problems
Stretched (1601); Material Deformation (2976)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/06/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The product has been received and the evaluation is pending.A follow up report will be submitted with investigation results once the evaluation has been completed.Section requested but not provided.
|
|
Event Description
|
It was reported from unit lakeside 55 that the primary tubing set was infusing via alaris pcu and pump module for less than 24 hours when the tubing set developed a balloon in the silicone segment.The customer later stated that there were no adverse effects caused to the patient from the event.
|
|
Manufacturer Narrative
|
The customer¿s report of ballooning was confirmed by visual inspection of the as-received sample.The set was visually inspected for kinks, incomplete bonding engagements, holes/tears in the tubing or damages to the components.Visual inspection observed a deformation at the top of the silicone pump segment indicating that it had previously ballooned.Further visual inspection under magnification found the walls of the silicone tubing segment to be concentric.Functional testing was not conducted.Previously investigated complaints for this same failure mode determined that the ballooning is caused by excess pressure within the silicone tubing segment.The root cause for the source of the excessive pressure is unknown.
|
|
Event Description
|
It was reported from unit lakeside 55, that the primary tubing set was infusing via an alaris point of care unit, (pcu) and pump module for less than (24) hours when the tubing set developed a "balloon" in the silicone segment, after the nurse heard the pump alarming.The customer later stated that there were no adverse effects caused to the adult patient from this event.
|
|
Search Alerts/Recalls
|