Model Number G447
Device Problems High impedance (1291); Low impedance (2285)
Patient Problems Cardiac Arrest (1762); Heart Failure (2206)
Event Date 08/23/2019
Event Type  malfunction  
Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
It was reported that this patient had cardiac arrest and received cardiopulmonary resuscitation (cpr).The patient was implanted with a left ventricular assist device (lvad) following the arrest.A review of the device data identified an out of range shocking impedance measurement greater than 200 ohms occurred one day after the lvad procedure.Additionally, pacing impedance measurements on the right ventricular (rv) channel have been trending lower since the procedure and there is recent stored episodes of noise on the shock channel.A boston scientific technical services consultant discussed the clinical observations with the caller who stated the root cause of the clinical observations has not been determined.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESONATE X4 CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key9638677
MDR Text Key176651328
Report Number2124215-2019-28746
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526589539
UDI-Public00802526589539
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S436
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 01/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/22/2021
Device Model NumberG447
Device Catalogue NumberG447
Device Lot Number208120
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/06/2019
Initial Date FDA Received01/28/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/06/2019
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Age52 YR
-
-