Model Number SGC0701
Device Problems Deformation Due to Compressive Stress (2889); Material Split, Cut or Torn (4008)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/19/2019
Event Type  malfunction  
Manufacturer Narrative
The device was received.Investigation is not yet complete.A follow up report will be submitted with all additional relevant information.The cds-ntw referenced was filed under mfr# 2024168-2020-00552-00.
 
Event Description
This is filed to report the torn soft tip observed on the returned steerable guide catheter (sgc) which was used with the first clip delivery system (cds).It was reported that this was a mitraclip procedure to treat grade 4 mixed etiology mitral regurgitation (mr).The sgc and the ntw cds were advanced to the left atrium via right groin access without issue.There was difficulty opening the clip, but the clip popped open and the clip had detached from the cds.It was decided to remove the cds with the clip.The clip was retracted to the sgc tip via the lock line.Left groin access was then obtained and a guide wire was advanced across the septum next to the sgc.A 14 french balloon dilatation catheter (bdc) was advanced to the septum and dilated the septum to allow for passage of the open clip.The bdc was removed and the sgc was retracted with the open clip across the septum into the right atrium and down to the groin.A new sgc and new cds were advanced to the left atrium via the left groin access.The mitral valve was treated and mr was reduced to grade 1 with the single clip.The atrial septal defect was closed with a septal occluder.The first sgc and clip were then fully removed from the patient via the right groin.Pressure was applied and the access sites were sutured.The patient is doing great.The following day, the patient was discharged home from the hospital with mr grade 1.No additional information was provided.
 
Manufacturer Narrative
The steerable guide catheter (sgc) was returned.The sgc soft tip was noted to be deformed during returned device analysis.The deformed sgc soft tip appears to be due to procedural circumstances.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot.Additionally, a review of the complaint history identified no other complaints reported from this lot.Based on the information reviewed, there is no indication of a product quality issue with respect to manufacture, design or labeling.H6: device code 4008 was removed and replaced with device code 2889.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 STEERABLE GUIDE CATHETER
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key9639834
MDR Text Key177379574
Report Number2024168-2020-00958
Device Sequence Number1
Product Code DRA
UDI-Device Identifier08717648231025
UDI-Public08717648231025
Combination Product (y/n)N
PMA/PMN Number
K190167
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 03/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/23/2020
Device Model NumberSGC0701
Device Catalogue NumberSGC0701
Device Lot Number91024U222
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/14/2020
Initial Date Manufacturer Received 01/14/2020
Initial Date FDA Received01/28/2020
Supplement Dates Manufacturer Received02/11/2020
Supplement Dates FDA Received03/04/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
CDS0701-NTW.
Patient Age68 YR
Patient Weight73
-
-