• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® SILASTIC® FOLEY CATHETER; SILASTIC FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® SILASTIC® FOLEY CATHETER; SILASTIC FOLEY CATHETER Back to Search Results
Model Number 33618
Device Problems Misassembled (1398); Failure to Align (2522)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.
 
Event Description
It was reported that the inflation cap was misaligned on one of the catheters.
 
Manufacturer Narrative
The reported event was confirmed.Seven unopened silastic catheters were returned.A third catheter was opened and it was found that the catheter's drainage funnel appeared to be poorly trimmed.The catheter was inflated with air and passively deflated with no issues.The same process was repeated with 10ccs of water and initially no water returned.The syringe was screwed further and the balloon passively deflated with no issues.Upon closer examination it was found that the catheter cap was misaligned.A potential root cause for this failure could be "dipping too shallow - incorrect fluid level in the slurry tank".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿to deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If necessary, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that the inflation cap was misaligned on one of the catheters.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® SILASTIC® FOLEY CATHETER
Type of Device
SILASTIC FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key9641306
MDR Text Key184139855
Report Number1018233-2020-00571
Device Sequence Number1
Product Code EZL
UDI-Device Identifier00801741020070
UDI-Public(01)00801741020070
Combination Product (y/n)N
PMA/PMN Number
K951103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Type of Report Initial,Followup
Report Date 03/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date05/31/2023
Device Model Number33618
Device Catalogue Number33618
Device Lot NumberNGDS0376
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/26/2019
Initial Date Manufacturer Received 01/07/2020
Initial Date FDA Received01/29/2020
Supplement Dates Manufacturer Received03/04/2020
Supplement Dates FDA Received03/09/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-