• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NEUROSURGERY STRATA II VALVE, SMALL, WITH BIOGLIDE; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROSURGERY STRATA II VALVE, SMALL, WITH BIOGLIDE; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS Back to Search Results
Model Number 92856
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/27/2020
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported there was a concern of valve obstruction after implantation.It was noted that there was no obstruction in the ventricle and peritoneal cavity catheters so only the valve was replaced with another one of the same model but from a different lot.The patient's status at the time of the report was alive-no injury.
 
Manufacturer Narrative
The returned valve was patent.It met the requirements for leakage, reflux, valve flux, siphon control and preimplantation testing.The valve did not meet the requirements for siphon and pressure flow testing.Biological and crystalline debris were observed inside the delta chamber.The instructions for use cautions, ¿the system may become occluded internally due to tissue fragments, blood clots, tumor cell aggregates, bacterial colonization or other debris.¿ dhr review performed on lot # e63448 did not indicate any finding correlated to the complaint.All valves are 100% tested at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRATA II VALVE, SMALL, WITH BIOGLIDE
Type of Device
SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
5290 california ave
irvine CA 92617
Manufacturer (Section G)
MEDTRONIC NEUROSURGERY
5290 california ave
irvine CA 92617
Manufacturer Contact
katcha taylor
9775 toledo way
irvine, CA 92618
9496801345
MDR Report Key9643656
MDR Text Key176851232
Report Number2021898-2020-00048
Device Sequence Number1
Product Code JXG
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K042465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number92856
Device Catalogue Number92856
Device Lot NumberE63448
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/28/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/27/2020
Initial Date FDA Received01/29/2020
Supplement Dates Manufacturer Received05/15/2020
Supplement Dates FDA Received05/15/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-