• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US UNKNOWN HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US UNKNOWN HIP FEMORAL STEM Back to Search Results
Catalog Number UNK HIP FEMORAL STEM
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Edema (1820); Foreign Body Reaction (1868); Necrosis (1971); Sepsis (2067); Osteolysis (2377); Test Result (2695); No Code Available (3191)
Event Date 11/27/2018
Event Type  Injury  
Manufacturer Narrative
(b)(4).Initial reporter occupation: lawyer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Asr medical records received.After review of medical records, the patient was revised to address a failed right total hip arthroplasty due to metallosis, elevated metal ions with sepsis.Intraoperatively, approximately 100 ml of abundant cloudy fluid poured out from the capsule.The capsule was consistent with adverse reaction to metal debris.The femoral component was stable but there was a necrotic, osteolytic lesion.The patient was noted to be about 4mm short preoperatively.The cup was noted to have 59 degrees of inclination with significant anteversion.However, the surgeon did not consider the cup to be malpositioned given the patient's severe spinopelvic mal-alignment.The surgeon elected to retain the asr xl cup and the stem.Doi: (b)(6) 2009; dor: (b)(6) 2018 (right hip).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.A worldwide lot specific complaint database search, or manufacturing record evaluation, was not possible as the required lot number was not provided.The information received will be retained for trend analysis, post market surveillance, or other events within the quality system.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN HIP FEMORAL STEM
Type of Device
HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9647886
MDR Text Key186841326
Report Number1818910-2020-03613
Device Sequence Number1
Product Code KXA
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup
Report Date 01/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberUNK HIP FEMORAL STEM
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/17/2020
Initial Date FDA Received01/30/2020
Supplement Dates Manufacturer Received04/22/2020
Supplement Dates FDA Received04/24/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age76 YR
-
-