The incident sample has been requested but to date has not been received for evaluation.If the sample is received, or if additional information pertinent to the incident is obtained a follow-up report will be submitted.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.
|
The device history record was reviewed and indicated that the product was released accomplishing all quality standards.One sample has been received for the evaluation.The sample was evaluated by pumping test.Upon evaluation, leaking from aff valve was observed.Therefore, the reported condition is confirmed.Based on the evaluation, incomplete aff valve assembly from the supplier caused the product leakage.The supplier team has determined that the root cause could be because of the machine might not have been received right maintenance.The detection does not have a documented periodic inspection of the sensor.As a containment action, refresh training and establish visual control to operator who conduct 100% screen before assembly has been created.Full review of current status of pistons and sensors for the machines will be performed through maintenance orders.Analysis of means (anom) inspection is currently implemented along with quality inspection.
|