Model Number SR-200-NPS
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Patient Problem/Medical Problem (2688); Vascular Dissection (3160)
Event Date 08/18/2017
Event Type  Injury  
Manufacturer Narrative
On 17dec2019 silk road medical contacted osb to inform them of medical device reports that were not reported within the mandatory reporting timeframe.It is noted here, per fda's request, that this mdr was identified as part of a complaint handling remediation effort associated with capa-(b)(4).The product in complaint was not returned to the manufacturer for analysis.A review of the manufacturing records for this device was completed and no issues were identified that could lead to the adverse event reported.Complaints will continue to be monitored for any trends.
 
Event Description
Patient underwent tcar procedure.The arterial sheath was advanced over the guidewire and went in smoothly under angiography, with wire never advancing, and was then sutured into place.Dissection noted at the tip of the arterial sheath when taking 2 views.Attempts were made to pull back arterial sheath to allow wire into true lumen.Original access point was closed, and the physician attempted the process again.Angio revealed dissection, but md did not feel he was in true lumen and did not wish to try and pass wires at this time.Decision was made to abandon transcarotid revascularization procedure, and convert treatment to cea.Patient did well and woke up post procedure neurologically intact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENROUTE TRANSCAROTID NEUROPROTECTION SYSTEM
Type of Device
TEMPORARY CAROTID CATHETER FOR EMBOLIC CAPTURE
Manufacturer (Section D)
SILK ROAD MEDICAL, INC.
1213 innsbruck drive
sunnyvale CA 94089
Manufacturer (Section G)
SILK ROAD MEDICAL, INC.
1213 innsbruck drive
sunnyvale CA 94089
Manufacturer Contact
1213 innsbruck drive
sunnyvale, CA 94089
4087209002
MDR Report Key9650068
MDR Text Key179312262
Report Number3014526664-2020-00010
Device Sequence Number1
Product Code NTE
UDI-Device Identifier00811311020744
UDI-Public(01)00811311020744(17)171029(10)200723
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K153485
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 01/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator No Information
Device Expiration Date10/29/2017
Device Model NumberSR-200-NPS
Device Catalogue NumberSR-200-NPS
Device Lot Number200723
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received01/30/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-