Model Number CD3371-40QC |
Device Problem
Fitting Problem (2183)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 01/17/2020 |
Event Type
malfunction
|
Event Description
|
During a unrelated lead revision procedure, there was a connection issues between the header of the device and the lead.The device was explanted and replaced to resolve the event and the patient was in stable condition throughout the event.
|
|
Manufacturer Narrative
|
The reported field event of set screw anomaly was verified in the lab.The df4 and the is-1 set screws were found to be completely backed out from the set screw threads as a result of unscrewing the set screws too far.When the screws were re-engaged with the connector block, the screw operated normally in affixing a test lead to the device.The issue was consistent with the procedure.Interrogation of the device revealed the device was above elective replacement indicator (eri) when received.The device pacing lead impedances and sensing were tested on the bench and they were found to be normal.
|
|
Event Description
|
During a unrelated lead revision procedure, the atrial lead was unable to be connected to the device due to a set screw anomaly.The device was explanted and replaced to resolve the event and the patient was in stable condition throughout the event.
|
|
Search Alerts/Recalls
|