• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 LUBRI-SIL® I.C. COMPLETE CARE® TEMPERATURE-SENSING URINE METER FOLEY TRAY; URINE DRAINAGE BAG (COMPLETE CARE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 LUBRI-SIL® I.C. COMPLETE CARE® TEMPERATURE-SENSING URINE METER FOLEY TRAY; URINE DRAINAGE BAG (COMPLETE CARE) Back to Search Results
Catalog Number 319214A
Device Problems Contamination /Decontamination Problem (2895); Device Contamination with Chemical or Other Material (2944)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Date 12/11/2019
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.
 
Event Description
It was reported that foreign material was found inside the meter.
 
Event Description
It was reported that foreign material was found inside the meter.
 
Manufacturer Narrative
The reported event was confirmed as manufacturing related.Visual evaluation of the returned sample noted one unopened (inside original packaging), meter drain bag with tray components and foley catheter.Visual inspection of the sample noted a loose yellow fibrous material inside the sealed meter of the bag.The material measured greater than 5.00 sq mm.This is out of specification per inspection procedure which states, "assemblies and packaging must be clean and free from loose or embedded foreign matter greater than 1/16¿ in length, and not to exceed an aggregate total of 0.6mm² in area, per the dirt estimation chart or other equivalent standard." although the reported event was confirmed, the root cause could not be determined.A potential root cause for this failure could be, ¿no follow up to cleaning procedure in the production areas.¿ the device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "[warnings] 1.Method for use (1) do not inflate the balloon in the urethra.[the urethra may be injured.] (2) do not pull the catheter hard.[the bladder/urethra may be injuryed.] 2.Applicable patients patients with delirium who might pull out catheter [when patient tugs at catheter unconsciously, the bladder and urethra may be damaged.] [contraindications] 1.Method for use: (1) do not reuse (2) do not resterilize (3) this device contains 10% povidine-iodine.For patients with past history of allergic hypersensitivity to povidone-iodine or iodine, consider using alternative disinfectants.(4) be careful that the catheter is not exposed to ointments, contract medium or oil-based lubricants (including vegetable oils such as olive oil, mineral oils such as white petrolatum and animal oils).[they may damage the device and may burst balloon.] (5) do not hold the device with forceps, etc.Avoid contact with any blades or sharp-edged instruments.[catheter damage may cause balloon rupture and accidental balloon removal or failure to deflate or remove the balloon.] directions for use: 1) clean the area around the external urethral meatus with the packaged cotton balls immersed in the antiseptics.2) lubricate the catheter shaft with the lubricant jelly.3) carefully insert the catheter into the urethral meatus.After the balloon advanced in the bladder, attach the needleless syringe, and gently infuse the specified volume of sterile water to inflate the balloon.4) pull the catheter slightly to seat the balloon at the level of the bladder neck.5) to deflate and remove the balloon, attach a needleless syringe to let sterile water in the balloon come out spontaneously through balloon deflation without aspiration.After balloon deflation, withdraw the catheter slowly while confirming that no abnormal resistance is encountered." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUBRI-SIL® I.C. COMPLETE CARE® TEMPERATURE-SENSING URINE METER FOLEY TRAY
Type of Device
URINE DRAINAGE BAG (COMPLETE CARE)
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key9654066
MDR Text Key183971530
Report Number1018233-2020-00701
Device Sequence Number1
Product Code MJC
Combination Product (y/n)N
PMA/PMN Number
K070582
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,other
Type of Report Initial,Followup
Report Date 03/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/31/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/28/2023
Device Catalogue Number319214A
Device Lot NumberMYDS5507
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/23/2020
Date Manufacturer Received03/20/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-