• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HUDSON MICRO MIST NEBULIZER W/ELONG; NEBULIZER (DIRECT PATIENT INTE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HUDSON MICRO MIST NEBULIZER W/ELONG; NEBULIZER (DIRECT PATIENT INTE Back to Search Results
Catalog Number 1885
Device Problem Gas/Air Leak (2946)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/26/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed and no relevant findings were identified.Without the device to evaluate, the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
 
Event Description
Customer reported the patient's oxygen saturation had decreased to 80% and the device was placed on the patient to improve oxygen saturation.The oxygen was set at 7l/min, but the patient's saturation did not change.The nurse heard a leak and found the mask tube was cracked.The mask was changed and the patient's oxygen saturation increased to 95%.Patient's condition reported to be "fine.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON MICRO MIST NEBULIZER W/ELONG
Type of Device
NEBULIZER (DIRECT PATIENT INTE
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
parque industrial finsa
nuevo laredo 88275
MX   88275
Manufacturer Contact
victoria sandlin
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key9659033
MDR Text Key189844285
Report Number3004365956-2020-00025
Device Sequence Number1
Product Code CAF
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K930525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date07/06/2023
Device Catalogue Number1885
Device Lot Number74G1800185
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/16/2020
Initial Date FDA Received02/03/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/02/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-