MAUDE Adverse Event Report: HOLOGIC, INC SELENIA DIMENSIONS MAMMOGRAPHY SYSTEM; FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOLOGIC, INC SELENIA DIMENSIONS MAMMOGRAPHY SYSTEM; FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM Back to Search Results
Model Number SDM-00001-2D
Device Problem Unintended System Motion (1430)
Patient Problem Bruise/Contusion (1754)
Event Date 01/13/2020
Event Type  malfunction  
Manufacturer Narrative
As of today the investigation is still in process and a follow up will be filed as needed.
 
Event Description
It was reported that while a patient was in compression the c-arm rotated.The customer reported the patient had a bruised breast the next day.A field engineer was dispatched to the site.
 
Event Description
It was determined that the rear rotation switch, lateral rotation switch, and power distribution drawer needed to be replaced.Once this was completed the system was working as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELENIA DIMENSIONS MAMMOGRAPHY SYSTEM
Type of Device
FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM
Manufacturer (Section D)
HOLOGIC, INC
36 & 37 apple ridge road
danbury CT 06810
MDR Report Key9661243
MDR Text Key178629357
Report Number1220984-2020-00015
Device Sequence Number1
Product Code MUE
Combination Product (y/n)N
PMA/PMN Number
P010025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Type of Report Initial,Followup
Report Date 01/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSDM-00001-2D
Device Catalogue NumberSDM-00001-2D
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/14/2020
Initial Date FDA Received02/03/2020
Supplement Dates Manufacturer Received04/15/2020
Supplement Dates FDA Received04/20/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-