Model Number CD3361-40C |
Device Problem
Fitting Problem (2183)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 01/22/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The results/method and conclusion codes along with investigation results will be provided in the final report.
|
|
Event Description
|
During a routine device exchange, the right atrial (ra) lead was not able to be connect into the header of the device.The device was replaced to resolve the event with no malfunction on the ra lead.The patient was stable with no consequences.
|
|
Manufacturer Narrative
|
The device was received with right atrial (ra) set screw fully engaged, blocking the insertion of ra lead.The ra set screw was disengaged with a lab torque driver; then test leads were able to fully insert into the device header.Device set-screws were able to be fully engaged with a lab torque driver and were revealed to function normally.The test leads were firmly secured.No anomalies were noted.
|
|