• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD HEMI HEAD 50MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD HEMI HEAD 50MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Model Number 74122550
Device Problems Insufficient Information (3190); Appropriate Term/Code Not Available (3191)
Patient Problems Pain (1994); Injury (2348)
Event Date 10/17/2019
Event Type  Injury  
Event Description
It was reported a right hip revision surgery due to persistent pain and extensive trunnionosis.
 
Manufacturer Narrative
It was reported that right hip revision surgery was performed.During the revision, the hemi head and modular sleeve removed.The echelon stem and acetabular cup remained implanted.As of today, the implanted devices, all of which were used in treatment, and additional information have been requested for this complaint but have not become available.A review of the complaint history for the hemi head, acetabular cup, modular sleeve and stem was performed using batch numbers in search of similar recurring reports for the products during their lifetimes.No other similar complaints were identified.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.All the released devices involved met manufacturing specifications at the time of production.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.No additional risks were identified as result of the reported event.The available medical documents were reviewed.The reported pain and intraoperative findings of extensive trunnionosis may be consistent with findings associated with metal debris.However, without the supporting lab/pathology results, imaging, and/or the analysis of the explanted components, the root cause of the reported pain and trunnionosis cannot be confirmed, and it cannot be concluded that the reported clinical reactions were associated with a mal-performance of the implant.The patient impact beyond the pain, revision, and expected transient post-op convalescence period cannot be determined.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMI HEAD 50MM
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house
spa park
leamington spa CV31 3HL
UK  CV31 3HL
MDR Report Key9671699
MDR Text Key177815589
Report Number3005975929-2020-00041
Device Sequence Number1
Product Code NXT
UDI-Device Identifier00885556071205
UDI-Public00885556071205
Combination Product (y/n)N
PMA/PMN Number
P040033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup
Report Date 04/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/05/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator No Information
Device Expiration Date10/29/2012
Device Model Number74122550
Device Catalogue Number74122550
Device Lot Number07KW14092
Date Manufacturer Received01/23/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ECH PRI STEM 71341016/ (B)(6).; MODULAR SLEEVE 74222100/ (B)(6).; ECH PRI STEM 71341016/ 00808946A; MODULAR SLEEVE 74222100/ 07LW14954
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age61 YR
-
-