The comet was returned for analysis: the tip, device shaft, sensor port and the coefficient values were examined for damage or any irregularities.The devices shaft showed three kinks located 38cm, 140cm from the tip and behind the occ handle collet.The tip showed a bend.There was peeled coating at the 140cm kink.The occ handle was connected to the ffr link for signal verification.The signal was not present as designed.The sensor port was inspected to verify that the sensor was connected to the fiber optic.It was noticed that the sensor was in the correct location in the sensor port.The wire was gently moved back and forth to see if the sensor would move.The sensor did not move which verifies that the fiber optics were connected to the sensor.The proximal end of the wire was inspected for any fiber optic cracks or damage and that is was polished correctly.The wire end showed no damage.The occ handle cap was loosened to remove the wire.There was no issue with removing the wire.The sensor port showed residue of body fluids.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.
|