• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® I.C. FOLEY CATHETER; FOLEY CATHETER (LATEX)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® I.C. FOLEY CATHETER; FOLEY CATHETER (LATEX) Back to Search Results
Model Number 0165SI16
Device Problems Deflation Problem (1149); Device Damaged Prior to Use (2284)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 01/14/2020
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.
 
Event Description
It was reported that a mixture of urine and sterile water was aspirated when the catheter balloon was deflated.
 
Event Description
It was reported that a mixture of urine and sterile water was aspirated when the catheter balloon was deflated.
 
Manufacturer Narrative
The reported event was unconfirmed since the problem could not be reproduced.The evaluation observed that the surface of the catheter had a change in color that appeared to have been in use for a long period.The catheter was able to inflate and deflate without any difficulty.However, the defect was unable to determine for returned sample due to the water that retrieve from catheter was not show any color changes.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿general used if there is no specific defect (1) do not reuse.(2) do not resterilize.(3) this device contains 10% povidone-iodine.For patients with past history of allergic hypersensitivity to povidone-iodine or iodine, consider using alternative disinfectants.(4) be careful that the catheter is not exposed to ointments, contrast medium or oil-based lubricants (including vegetable oils such as olive oil, mineral oils such as white petrolatum and animal oils).[they may damage the device and may burst balloon.] (5) do not hold the device with forceps, etc.Avoid contact with any blades or sharp-edged instruments.[catheter damage may cause balloon rupture and accidental balloon removal or failure to deflate or remove the balloon.] 2.Applicable patients (1) patients who are or have been allergic to natural rubber latex (2) patients with known allergy to silver coated catheter".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® I.C. FOLEY CATHETER
Type of Device
FOLEY CATHETER (LATEX)
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key9672729
MDR Text Key182183960
Report Number1018233-2020-00825
Device Sequence Number1
Product Code MJC
UDI-Device Identifier00801741016752
UDI-Public(01)00801741016752
Combination Product (y/n)N
PMA/PMN Number
K040658
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,other
Type of Report Initial,Followup
Report Date 03/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/05/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/01/2021
Device Model Number0165SI16
Device Catalogue Number0165SI16
Device Lot NumberMYCRBB08
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/22/2020
Date Manufacturer Received03/03/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-