The initial reporter stated they received discrepant low results for a total of 22 samples from 19 different patients tested with the elecsys ft4 ii assay on a cobas 8000 e 801 module.The reporter alleges that the ft4 measurements of the samples are low, compared with normal tsh values and ft4 measurements obtained with a siemens centaur xp analyzer.The patients had no clinical symptoms.Some samples were tested on two e 801 systems, with results from the first being labeled with "e1" and results from the second labeled with "e2".For results without "e1" or "e2" labels, it is unknown which e 801 system was used for testing.For samples 1-19, the e 801 ft4 results were reported outside of the laboratory.For samples 20-22, no incorrect results from these samples were reported outside of the laboratory.One of the e 801 analyzers is serial number (b)(4), but it is not known which of the two analyzers this serial number applies to.The serial number of the other e 801 analyzer was requested, but not provided.
|