Product analysis: a second device was returned for this event, and follow-up confirmed the device was related to this event.This everflex entrust device was received to the medtronic investigation lab for evaluation.The device was labelled as being associated with this complaint.The everflex entrust was returned inside the previously opened everflex entrust box and pouch which both indicated lot a840545.It was discovered a 0.014" guidewire was loaded and a non-medtronic introducer was returned.The deployment system was returned with a 0.014"" guidewire stuck loaded in the catheter.A non-medtronic sheath was returned with a suspected everflex stent protruding out from the distal end.The deployment system was inspected.It was observed the locking pin was removed and not included with the returned contents.The handle assembly showed inner assembly exposed between the right and the left handle assemblies.Approximately 22cm of the 0.014" gw was exposed outside of the distal end of the catheter shaft.A bend to the everflex entrust catheter shaft was noted at approximately 65cm from the distal tip.Direct light applied to the distal end of the catheter shaft found the stent was not loaded the catheter.The pusher was identified approximately 10cm from the distal tip of the catheter, which indicates the stent did not deploy as intended.The gw was unable to be removed from the deployment system due to encountered resistance.The handle assembly was cracked open.It was found the inner assembly was folded over itself and showed buckling.The pull cable detached from the silver outer bond.The area of the detachment shows the silver outer frayed out at the bond area which suggests exposure to excessive tensile force.The detached end of the pull cable remained intact.The non-medtronic introducer was inspected and observed a fractured stent was protruding out from the distal rim of the introducer by approximately 1cm.The tantalum markers at the end of the stent were not present, which indicates the stent fractured/broke off.The exposed stent showed biological material adhering to the struts.The exposed struts were bent in various directions.If information is provided in the future, a supplemental report will be issued.
|