The reported event was confirmed.Visual evaluation of the photo sample noted one opened (no original packaging), irrigation bulb syringe.It was noted that the syringe seemed to have a white band on it that was likely added by the user and a small greenish foreign material within the barrel, close to the tip.Although the foreign material could not be measured, as this was a photo sample, the foreign material size seemed to both greatly exceed an area of 0.6 sq mm and exceed a length of 1/16" and did not meet the specification "loose or embedded foreign matter greater than 0.6mm2 or 1/16¿ in length was not permitted.(3 particles maximum per side or surface).Although the reported event was confirmed, the root cause could not be determined.A potential root cause for this failure could be no follow up to the production areas cleaning procedure.The lot number is unknown; therefore, the device history record could not be reviewed.A labeling review was not performed because labelling could not have prevented the reported failure.
|