Product analysis was completed.The failure to program was not duplicated.A results of diagnostic testing indicated the device was operating properly.With the exception of parameters related to the reed switch, results of the final electrical test were successful.Functional bench testing demonstrate that consistent magnet activations are attainable.With the exception of the noted test-related issue, there were no adverse functional, mechanical, or visual issues identified with the returned generator.The error code 254 seen is an expected output for a generator that is disabled (in ¿stop¿ state) when interrogated with v1.6 software.As a functional check at the labeling interrogation step in manufacturing, the devices are interrogated.This interrogation transmits a ¿run¿ command and brings the devices out of the ¿stop¿ state.If this interrogation is missed in manufacturing, the m1000 device will be distributed in a ¿stop¿ state.If the m1000 generator is then interrogated by v1.6 programming software, the user will be warned that the generator is disabled due to a reason ¿254¿ and is not delivering therapy.This is not the case with previous versions of the software, in which the generator will enable upon interrogation and no message will display to the user.Note that this is not a software or generator error, and that the ¿stop¿ state is an acceptable device state.
|