• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT QUADRA ASSURA MP ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT QUADRA ASSURA MP ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3371-40QC
Device Problem Inappropriate or Unexpected Reset (2959)
Patient Problem Undesired Nerve Stimulation (1980)
Event Date 01/23/2020
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
The presented in clinic following a mri.The device was not programmed to mri settings prior to the mri which resulted in the device entering hardware backup vvi mode with no defibrillation function.The patient also reported phrenic nerve stimulation that began during the mri.Technical support was contacted and the device was reprogrammed to resolve the event.The patient was stable and will continued to be monitored.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ABBOTT
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ABBOTT
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
MDR Report Key9682193
MDR Text Key178171350
Report Number2938836-2020-00898
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeEZ
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 02/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2021
Device Model NumberCD3371-40QC
Device Lot NumberP000076383
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/23/2020
Initial Date FDA Received02/07/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/26/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age76 YR
Patient Weight80
-
-