• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC PORTEX GENERAL ANESTHESIA CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC PORTEX GENERAL ANESTHESIA CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE) Back to Search Results
Catalog Number C37101307J
Device Problem Gas/Air Leak (2946)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Device evaluation in progress.
 
Event Description
It was reported that during a pre-use check, the customer noticed that air was leaking from the anesthesia circuit.No patient injury or complications were reported in relation to this event.
 
Manufacturer Narrative
One anesthesia circuit was returned for evaluation.Visual inspection of the device found it to be in good physical condition.Device underwent leak testing, confirming the reported customer complaint.The problem source has been determined to be manufacturing.Process was reviewed; production personnel performs a 100 percent leak test for adult expandable circuits.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX GENERAL ANESTHESIA CIRCUITS
Type of Device
CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC
6000 nathan lane north
minneapolis MN 55442
MDR Report Key9683501
MDR Text Key178215677
Report Number3012307300-2019-06135
Device Sequence Number1
Product Code CAI
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/07/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date08/14/2022
Device Catalogue NumberC37101307J
Device Lot Number3855885
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/15/2020
Date Manufacturer Received02/25/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-