Catalog Number SP-101
Device Problems Patient-Device Incompatibility (2682); Compatibility Problem (2960)
Patient Problem Granuloma (1876)
Event Date 02/06/2020
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Physician used venaseal occluding device to treat the patient's entire great saphenous vein (gsv).Ifu was followed.A guide wire was used for the insertion of the catheter.It was reported that at patient follow up approximately 7 weeks post procedure, a granuloma was observed.The granuloma has been removed in a new procedure with venaseal.There was no further patient injury reported.
 
Manufacturer Narrative
Additional information: during index procedure, local anesthesia was used and compression was performed with the transducer.The catheter tip was located 5cm caudal to the saphenofemoral junction (sfj) prior to initial delivery of adhesive.The venaseal procedure was completed without issue and the vein closed.The granuloma was removed through a small incision in a separate procedure.No additional venaseal treatment was performed as was originally reported.The patient is doing well.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Additional information: the pathology report confirms the presence of non-malignant fibrinoid material in the excised granuloma.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENA SEAL CLOSURE SYSTEM
Type of Device
AGENT, OCCLUDING, VASCULAR, PERMANENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key9686205
MDR Text Key178320976
Report Number9612164-2020-00633
Device Sequence Number1
Product Code PJQ
Combination Product (y/n)N
PMA/PMN Number
P140018
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 05/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/31/2021
Device Catalogue NumberSP-101
Device Lot Number55792
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/06/2020
Initial Date FDA Received02/10/2020
Supplement Dates Manufacturer Received02/06/2020
02/19/2020
03/10/2020
04/29/2020
Supplement Dates FDA Received02/12/2020
02/24/2020
03/12/2020
05/08/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age56 YR
Patient Weight96
-
-