• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD ENDO CLIP III; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD ENDO CLIP III; CLIP, IMPLANTABLE Back to Search Results
Model Number 176630
Device Problem Fail-Safe Problem (2936)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/12/2019
Event Type  malfunction  
Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The instrument was received partially applied.No visual abnormalities were observed with the instrument.The instrument was cycled and the clip did not load, the pusher bar was observed not functioning properly.A review of the device history record indicates this product was released meeting all quality release specifications at the time of manufacture.However, a manufacturing fault was identified during product analysis.The root cause of the observed condition was determined to be a result of a manufacturing activity.The trip lever out of position was preventing the clips from loading properly.Improvements have been initiated to mitigate this condition.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during laparoscopic cholecystectomy procedure, the clips were able to load properly into the jaws, but when the surgeon attempted to fire the first clip it did not fire.The surgeon was not able to squeeze the handle and once the 2nd clip was loaded the jaws locked and the device could not be fired.The surgeon then used another device to resolve the issue in order to complete the case.There was no patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO CLIP III
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
Manufacturer (Section G)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO   0101
Manufacturer Contact
lisa hernandez
60 middeltown ave
north haven, CT 06473
2034925563
MDR Report Key9686629
MDR Text Key178347986
Report Number9612501-2020-00242
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10884521057852
UDI-Public10884521057852
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K100242
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 02/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2022
Device Model Number176630
Device Catalogue Number176630
Device Lot NumberJ9E1646Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/18/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/04/2020
Initial Date FDA Received02/10/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/14/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-