• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES HEMOSPHERE INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES HEMOSPHERE INSTRUMENT Back to Search Results
Model Number HEM1
Device Problems Incorrect Measurement (1383); Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 01/21/2020
Event Type  malfunction  
Manufacturer Narrative
The product is expected to be returned for product evaluation; but has not yet arrived.When the evaluation findings are available a supplemental submission will be sent.The device history record review was completed and all manufacturing inspections passed with no non-conformance's.The record of servicing has been reviewed and there is no previous related record.The udi number is (b)(4).An mdr for all 3 products involved will be submitted.The submission numbers will be provided in a supplemental report.
 
Event Description
It was reported by the clinicians that there were inaccurate svr, cardiac index and continuous cardiac index readings during patient monitoring.The hemisphere instrument (hem1), swan ganz module (sgm) and 70cc2 cable were involved and could not be ruled out as contributors to the issue.The svr reading was between 2900-3500.The ci reading was oscillating between 0.1 to 1.1 and the cci reading was 0.9.They were able to recall the oximetry data without any issues.The patient was administered 1500l of fluid.They powered down the hemisphere system and it was powered back up and then all components worked appropriately.The patient demographic information has been requested but not received.There was no patient harm or injury.
 
Manufacturer Narrative
The product has arrived for evaluation and the results are pending.When the evaluation findings are available a supplemental submission will be sent.The submission numbers for the other products involved in this event are: 2015691-2020-10464 and 2015691-2020-10465.
 
Manufacturer Narrative
One hemosphere hem1 instrument was returned for product evaluation.The three products involved in the event were connected for testing; hem1 instrument, sg module and the 70cc2 cable.The system verification test was performed and the testing was left to run for over three hours.The co values remained within appropriate parameters.A visual inspection was performed and there was no physical damage that was identified.There was no defect found.The reported issue could not be confirmed by product evaluation.With any hemodynamic monitoring readings can change quickly and dramatically.Clinicians are trained to evaluate the entire clinical presentation of the patient in order to make decisions.In addition, these devices are used by highly trained clinicians experienced in assessing and mitigating any hazards that arise.It could not be confirmed if user or procedural factors contributed to the stated event.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as part of the monthly review.
 
Manufacturer Narrative
Reference capa-20-00141.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMOSPHERE INSTRUMENT
Type of Device
HEMOSPHERE INSTRUMENT
Manufacturer (Section D)
EDWARDS LIFESCIENCES
one edwards way
irvine CA 92614
MDR Report Key9689278
MDR Text Key194478938
Report Number2015691-2020-10463
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
PMA/PMN Number
K163381
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup,Followup
Report Date 01/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/10/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/21/2023
Device Model NumberHEM1
Device Catalogue NumberHEM1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/26/2020
Date Manufacturer Received07/23/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age69 YR
Patient Weight85
-
-