• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION EXTENSION, 30CM; SCS EXTENSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION EXTENSION, 30CM; SCS EXTENSION Back to Search Results
Model Number 3383
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Irritation (1941); Pain (1994)
Event Date 12/01/2019
Event Type  Injury  
Manufacturer Narrative
Date of event is estimated.Based on the information provided a device problem was not identified, as a result a conclusive cause for the reported patient effect, and the relationship to the product, if any, cannot be determined.
 
Event Description
It was reported that the patient experienced irritation/pain at the extension site due to mechanical friction.As such, surgical intervention tool place on (b)(6) 2020 wherein the extension was explanted and replaced with a new extension addressing the issue.
 
Manufacturer Narrative
The reported event for discomfort at the extension site was not confirmed.The extension was received incomplete with only the terminal end lead segment being returned.The extension was cut due to the explant procedure.Full functional test could not be performed due to being incomplete.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTENSION, 30CM
Type of Device
SCS EXTENSION
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer Contact
andrea deitz
6901 preston road
plano, TX 75024
9723098000
MDR Report Key9693838
MDR Text Key178570829
Report Number1627487-2020-01467
Device Sequence Number1
Product Code GZB
UDI-Device Identifier05415067017918
UDI-Public05415067017918
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K960728
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/11/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date03/31/2017
Device Model Number3383
Device Catalogue Number3383
Device Lot Number5060519
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/21/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/02/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/22/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-