Model Number N/A |
Device Problems
Positioning Failure (1158); Difficult to Insert (1316)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 01/28/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Item #:unknown, unknown head lot #:unknown, item #:unknown, unknown stem lot #:unknown, item #:unknown, unknown cup lot #:unknown.Customer has indicated that the product will be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It has been reported that the liner would not sit down flush.No further information is available.
|
|
Manufacturer Narrative
|
(b)(4).This follow-up report is being submitted to relay additional information.Updated: b4, b5, d4, d10, g4, h2, h3, h4, h6 reported event was confirmed by visual examination of device.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.One g7 freedom const e1 lnr 36mm h was returned and evaluated.Upon visual inspection there is damage to the locking feature.There is also scratching on the outer radius.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|