MAUDE Adverse Event Report: MAQUET CARDIOVASCULAR LLC VASOVIEW HEMOPRO 2; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOVASCULAR LLC VASOVIEW HEMOPRO 2; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number VASOVIEW HEMOPRO 2
Device Problem Improper Flow or Infusion (2954)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/09/2020
Event Type  malfunction  
Manufacturer Narrative
Trackwise id # (b)(4).A lot history record review was completed for the reported product lot number.There were no ncmr¿s for the reported lot number.The device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.
 
Event Description
The hospital reported that during an endoscopic vein harvesting procedure using vasoview hemopro 2, no proper tunnel could be obtain during harvesting of the vein.All the necessary steps were taken to ensure that there was gas and no gas leaking but still no tunnel was obtained.A replacement device was used to complete the procedure.The hospital did not report any patient effects.
 
Manufacturer Narrative
Trackwise # (b)(4).The device was returned to the factory for evaluation on 01/28/2020.An investigation was conducted on 02/25/2020.A visual inspection was conducted.Signs of clinical use and evidence blood was observed on the btt.The btt was observed to be intact, no visual defects were observed on the btt.The harvesting device was observed to be intact, no visual defects were observed.The cannula was observed to be intact, no visual defects were observed.The insufflation tube and the water tube were observed to be intact.The c-ring appears to have been slightly tainted with blood as it appears to be reddish in color.The c-ring has no visible defects.Air flow on btt was tested on the btt insufflation port and air flow was not obstructed.The device was evaluated for the presence or absence of air flow through the distal insufflation tube using a cannula with the help of calibrated uson.A reference endoscope was inserted into a reference vv7 cannula and an air supply was connected to the distal insufflation tubing luer fitting.Air was passed through the insufflation port and was observed to flow freely through the distal port.To verify this, a pouch was sealed over the distal tip of the cannula.Air was passed through the cannula and the pouch was inflated.The air supply was stopped and the pouch stayed inflated.When gentle pressure was applied to the inflated pouch, the pouch deflated slightly, the air was turned back on and the pouch re-inflated.The test was then repeated for the reported cannula.The reference endoscope was inserted into the complaint device and evaluated for air flow.The device passed the air flow test, the co2 insufflation path on the complaint unit was open and unobstructed.The values displayed were within specified acceptable range which is 2198 sccm.Based upon the returned condition of the device and the results of the investigation, the reported failure "improper flow or infusion" was not confirmed.
 
Event Description
The hospital reported that during an endoscopic vein harvesting procedure using vasoview hemopro 2, no proper tunnel could be obtain during harvesting of the vein.All the necessary steps were taken to ensure that there was gas and no gas leaking but still no tunnel was obtained.A replacement device was used to complete the procedure.The hospital did not report any patient effects.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VASOVIEW HEMOPRO 2
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ 07470
MDR Report Key9697078
MDR Text Key189613098
Report Number2242352-2020-00136
Device Sequence Number1
Product Code GEI
Combination Product (y/n)N
PMA/PMN Number
K101274
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 02/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/08/2021
Device Model NumberVASOVIEW HEMOPRO 2
Device Catalogue NumberVH-4000
Device Lot Number25147894
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/28/2020
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 01/20/2020
Initial Date FDA Received02/11/2020
Supplement Dates Manufacturer Received02/25/2020
Supplement Dates FDA Received02/27/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-