Catalog Number 2400L |
Device Problem
Break (1069)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 01/17/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that a load cell module cable was kinked up and wires were exposed.This was identified during setup.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
H10: the device was received for evaluation.A visual inspection was performed which observed the scale cable assembly has a puncture through outer insulation but no physical damage to internal wires.A functional testing was performed with no issues noted.The reported condition was verified.The cause of the condition was end user improperly routing the scale cable assembly underneath main module when setting up and causing the main module to pinch through outer insulation of load cell cable assembly.The scale cable assembly was replaced during the service refurbishment process.A service history review revealed no indication that the parts replaced caused or contributed to the reported event.Should additional relevant information become available, a supplemental report will be submitted.
|
|