• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION EM 2400, LOAD CELL MODULE; SYSTEM/DEVICE, PHARMACY COMPOUNDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION EM 2400, LOAD CELL MODULE; SYSTEM/DEVICE, PHARMACY COMPOUNDING Back to Search Results
Catalog Number 2400L
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Date 01/17/2020
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a load cell module cable was kinked up and wires were exposed.This was identified during setup.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
H10: the device was received for evaluation.A visual inspection was performed which observed the scale cable assembly has a puncture through outer insulation but no physical damage to internal wires.A functional testing was performed with no issues noted.The reported condition was verified.The cause of the condition was end user improperly routing the scale cable assembly underneath main module when setting up and causing the main module to pinch through outer insulation of load cell cable assembly.The scale cable assembly was replaced during the service refurbishment process.A service history review revealed no indication that the parts replaced caused or contributed to the reported event.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EM 2400, LOAD CELL MODULE
Type of Device
SYSTEM/DEVICE, PHARMACY COMPOUNDING
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key9701792
MDR Text Key179233761
Report Number1416980-2020-00628
Device Sequence Number1
Product Code NEP
Combination Product (y/n)N
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 03/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/12/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2400L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/30/2020
Date Manufacturer Received02/19/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-