Catalog Number UE514010100 |
Device Problem
Device Emits Odor (1425)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 02/07/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Once the investigation is completed, a follow-up/final report will be submitted.Product evaluated by external contractor.
|
|
Event Description
|
It was reported that after surgery the unit had a burning smell and did not result in harm or delay.There was no smoke or spark.The cart did shut down with an over temp error.The new vacuum pump that was installed two days prior caused the burning smell.No adverse events were reported as a result of this malfunction.
|
|
Event Description
|
No additional event information is available.
|
|
Manufacturer Narrative
|
This event has been recorded under zimmer biomet complaint number (b)(4).Upon receipt of additional information, it was determined that the device in question did not cause or contribute to the reported event.This file was submitted in error and should be voided.
|
|