Related manufacturer reference number: 2017865-2020-02146, 2017865-2020-02148.It was reported that the patient presented in the hospital with diaphragmatic stimulation.Upon interrogation, the left ventricular lead exhibited high capture thresholds.A chest x-ray performed on 12 aug 2019 revealed left ventricular lead dislodgement, no lead slack on the right atrial lead, and sub-optimal right ventricular lead placement for high voltage therapy.The physician explanted and replaced the leads.The patient was discharged post-procedure.
|
A complete lead was returned for analysis in one piece measuring 86.1 cm.Visual and x-ray examination found procedural damage in the form of insulation cuts and no other anomalies.No conductor fractures or internal shorts were noted.The reported events of dislodgement, high capture thresholds, and diaphragmatic stimulation were not confirmed.
|