Upon receipt, the device was interrogated on a programmer and normal communication was established.Programmer printouts were obtained, and the device was decontaminated.After decontamination, a visual inspection of the device and torque wrench was performed and there was body fluid observed in the header of the device.The device header was inspected further under magnification and body fluid was observed in the right ventricular (rv) and left ventricular (lv) bores.The is4 and df4 septa were noted to be slightly damaged, which is consistent with procedural damage when multiple attempts to secure or disconnect the leads using a torque driver.This damage may result in body fluid contaminating the device header.The device was connected to test leads and resistor loads, and the high voltage lead impedance and pacing lead impedance were measured, and no anomalies were found.The device sensing was also tested, and the device was found to be normal.The reported event of body fluid in the header was confirmed, however the issue was found to be consistent with procedural damage.
|