It was reported that during a planned prophylactic generator replacement turned lead revision due to lead fracture (as reported in mfr report # 1644487-2020-00286), output current on the generator was unexpectedly 0 ma.Preoperatively, settings and diagnostics were within normal limits; however, when the attending representative tested the device after the lead fracture was observed the impedance was okay and the output current was 0 ma.No settings were changed.However, it is believed possible that diagnostics weren't run, given that the impedance value didn't change after a complete lead break was observed.Reportedly, no electrosurgical tools were used during the procedure.The explanted generator was discarded.No further relevant information has been received to date.
|