One catheter with attached bd 1.0 ml syringe was returned for evaluation.The balloon inflated clear, concentric and remained inflated for more than 5 minutes without leakage.The balloon deflated within 1 second without a syringe attached.The specification for balloon deflation without a syringe attached is 4 seconds.All through lumens were patent without any leakage or occlusion.No visible damage was observed from the catheter body or returned syringe.Balloon inflation test was performed using returned syringe with 0.8 cc air by holding the balloon under water for 5 minutes.Visual examination was performed under microscope at 10x magnification and with the unaided eye.Customer report of balloon deflation issue could not be confirmed during the analysis, as the device responded appropriately during functional testing.There was no evidence of a manufacturing nonconformance.No further actions will be taken at this time.Invasive procedures involve some patient risks.Although serious complications are relatively uncommon, the physician is advised, before deciding to insert or use the catheter, to consider the potential benefits in relation to the possible complications.The techniques for insertion, methods of using the catheter to obtain patient data information, and the occurrence of complications is well described in the literature.It is common clinical practice to check balloon integrity by inflating it to the recommended volume in order to detect any asymmetry, leakage condition or deflation issue before use of the catheter.In this event, there were no patient complications noted.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
|