Model Number 37800 |
Device Problem
High impedance (1291)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 02/07/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a healthcare professional (hcp) regarding a patient with an implantable neurostimulator (ins).It was reported that they ran impedances and found the values to be high.The hcp stated they programmed the patient using c+ 2- to avoid the high impedances and sent the patient for x-rays.The following values were provided: 2c 3703c >2000023 20000.No patient complications were reported as a result of this event.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received from the hcp.It was reported that the device was surgically replaced/revised.No patient complications were reported as a result of this event.
|
|
Manufacturer Narrative
|
Product id: 4351-35, serial# (b)(4), implanted: (b)(6) 2014, product type: lead; product id: 4351-35, serial# (b)(4), implanted: (b)(6) 2014, product type: lead.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information received from a healthcare provider (hcp).It was reported that the lead was surgically replaced and the patient is doing great.No further complications were reported.
|
|