• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065751763
Device Problem Complete Blockage (1094)
Patient Problem Eye Burn (2523)
Event Type  Injury  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported that there was a clogged tip, cloudiness from the tip at the beginning of ultrasound and the patient experienced a corneal burn.Additional information has been requested.
 
Manufacturer Narrative
Corneal thermal injuries are typically related to excessive heat generated by the phaco tip due to insufficient aspiration flow, extended energy application, or combination of both.The operator¿s manual states: appropriate use of vision system parameters and accessories is important for successful procedures.Use of low vacuum limits, low flow rates, low bottle heights, high power settings, extended power usage, power usage during occlusion conditions (beeping tones), failure to sufficiently aspirate viscoelastic prior to using power, excessively tight incisions, and combinations of the above actions may result in significant temperature increases at incision site and inside the eye, and lead to severe thermal eye tissue damage.Good clinical practice dictates testing for adequate irrigation, aspiration flow, reflux, and operation as applicable for each handpiece prior to entering eye.Ensure that the tubings are not occluded during any phase of operation.Use of a phaco handpiece in the absence of irrigation flow and/or in the presence of reduced or lost aspiration flow and/or sideways orientation of the tips can cause excessive heating and potential thermal injury to adjacent eye tissues.Directing energy toward non-lens material, such as iris or capsule, may cause mechanical and/or thermal tissue damage.The vision system is designed to cool the phaco tip during use as aspirated fluid flows through the tip lumen.Overheating of the phaco tip, however, may occur due to extended application of ultrasonic energy or compromised aspiration flow through the phaco tip.Reduced fluid flow through the phaco tip may be caused by phaco tip re-use, tip clogging by nuclear material, kinked tubing, inadequate flow and vacuum settings, or obstruction by ophthalmic viscoelastic device (ovd).When the phaco tip is occluded, infusion will cease, reducing the cooling effect of the tip.Occlusion tones (intermittent beeping tones during occlusion) alert the user, indicating that the vacuum is near or at its preset limit, and aspiration flow is reduced or stopped.The surgeon must recognize the occlusion tones and manually stop the ultrasound mode in order to prevent a rapid temperature increase.Corneal burn is an issue that is occasionally reported with cataract surgery.According to the pennsylvania patient safety advisory abstract: preventing corneal burns during phacoemulsification, march 2010, vol.7, no.1: 23-25, most corneal burns can be traced to issues related to surgical technique and not to malfunctioning equipment.The customer called the company technical support specialist (tss) asking about the possible causes of a clogged tip, and the cloudiness from the tip at the beginning of phaco power sculpting.The tss explained about the warning occlusion tones from the system.The customer has not requested service for the system.The system manufacturing device history record (dhr) was reviewed.Based on qa assessment, the product met specifications at the time of release.With no additional, related information provided, the customer reported event could not be confirmed.The root cause cannot be determined conclusively.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key9730333
MDR Text Key180098060
Report Number2028159-2020-00115
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
PMA/PMN Number
K121555
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Remedial Action Other
Type of Report Initial,Followup
Report Date 04/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/19/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065751763
Was Device Available for Evaluation? No
Date Manufacturer Received03/16/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
PHACO TIP, TURBOSONIC ABS; PHACO TIP, TURBOSONIC ABS
Patient Outcome(s) Other;
-
-