(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: according to the information provided, it was reported that during an unknown procedure the expressew iii suture passer w/o hook and the clever hook - left were broken.The jaws of the devices are not opening and closing.The complaint device was received and evaluated.Visual inspection was performed and not significant damage was observed functional testing revealed difficulties operating the device.There was resistance as trying to close and open the jaws, the jaw only move as manual mode, the lower jaw was loose and it was not allow to hold, therefore the complaint was confirmed.The possible root cause for the reported failure can be attributed to use normal wear and tear of the device as the device is reusable, however; this cannot be conclusive determined. a manufacturing record evaluation was performed for the finished device [09g3] number, and no non-conformances were identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.(b)(4).
|