Model Number 134044
Device Problem Mechanics Altered (2984)
Patient Problem No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, the handle did not move well after clipping, and the loading was slow.
 
Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The instrument was received partially applied with five remaining clips.Visual inspection of the instrument revealed no abnormalities.Functionally, the instrument was fired once in air and proper clip formation was confirmed.The remaining clips were then fired on test media with proper formation.The jaw and handle moved smoothly through the firing cycle and returned to the open position and each remaining clip loaded properly in the jaw.When the cartridge was empty, the interlock engaged and prevented the jaw from approximating.Note that the information for use brochure which accompanies each product shipment cautions the user as follows: ensure that the handles are squeezed together firmly as far as they will go.Failure to squeeze the handles completely can result in clip malformation and possible bleeding or leakage.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PREMIUM SURGICLIP
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer (Section G)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer Contact
lisa hernandez
60 middletown ave.
north haven, CT 06473
2034925563
MDR Report Key9743021
MDR Text Key180514360
Report Number2647580-2020-00672
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10884521057531
UDI-Public10884521057531
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
K853650
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/23/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2023
Device Model Number134044
Device Catalogue Number134044
Device Lot NumberP8C1529CX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/12/2020
Is the Reporter a Health Professional? No
Date Manufacturer Received04/06/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/27/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-