Related manufacturer reference number: 2030404-2020-00012, 3008452825-2020-00112, 2182269-2020-00020, 2182269-2020-00021, 3008452825-2020-00113, 2030404-2020-00013, 3005334138-2020-00072.During a pulmonary vein isolation ablation procedure, a pericardial effusion occurred.After the procedure was completed, the patient became hypotensive, so an ultrasound was performed, confirming a pericardial effusion.A pericardiocentesis was performed to stabilize the patient.There were no performance issues with any abbott device.
|
One 6f, decapolar, medium sweep, livewire ep catheter was received for evaluation.No visual or function issues were noted.Electrodes 1-10 met specifications of acceptable resistance values with no open or short circuits detected.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported cardiac perforation cannot be conclusively determined.Per the ifu, cardiac perforation is a known risk during the use of this device.
|