(b)(4).The customer returned one unit ae05ml autoend05 ml for investigation.The returned sample was visually examined with and without mag nification.Visual examination of the returned device revealed that the sample appears used as there is biological material present on the device.Reference file anp1900075600 for investigation photos.Functional inspection was performed on the returned sample by attempting to engage the trigger using hand pressure.Upon engagement of the trigger, an audible ratchet sound could be heard indicating that the internal ratchet ears are intact.The first clip was unable to load properly.Some abnormal resistance was felt during the trigger pull.The sample was disassembled to inspect the internal components.It was found that the clips were out of position and stacking on one another in the channel.The sample was received with 5 clips remaining in the channel, indicating that 10 clips were fired by the end user.The clip stacking prevented the clips from loading properly into the jaws.It could not be determined exactly how or when the clips became out of position.A capa has been opened to further investigate this issue.Reference attached file anp1900075600 for investigation photos.The ifu for this product, l06072, was reviewed as a part of this complaint investigation.The ifu for this product states, "mishandling of appliers may result in improper load and/or closure of the ligating clip." the clip stacking prevented the clips from loading properly into the jaws.It could not be determined exactly how or when the clips became out of position.A capa has been opened to further investigate this issue.The reported complaint of "misfire" was confirmed based upon the sample received.One device was returned with 5 clips remaining in the channel, indicating that 10 clips were fired by the end user.Upon functional inspection, the first clip was unable to load properly.The sample was disassembled and it was found that the clips were out of position and stacking on one another.The clip stacking prevented the clips from loading properly into the jaws.Although the reported complaint issue was confirmed based on functional testing, it could not be determined exactly how or when the clips became out of position.A capa has been opened to further investigate this issue.
|