• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE Back to Search Results
Model Number EL5ML
Device Problem Physical Resistance/Sticking (4012)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/01/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # t94u28.Investigation summary: the analysis results found that the el5ml device was received with no damage in the external components.Upon cycling, the instrument was noted to be empty and locked out.The instrument is designed to lock out after all the clips have been fired.A potential cause of the customer reported experience is the firing of all of the clips and the instrument "will not fire" (activation of the lock out mechanism).The instrument has an orange indicator that appears on the top of the handle as a reference for the user as to the quantity of clips remaining.The event described could not be confirmed as the device was returned empty.In addition, a photo was received for review.Upon visual inspection of the photo, the following was observed: the photo shows a device from jaws area and no anomalies could be observed.Based on the photo alone, the event described cannot be confirmed.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances related to the reported complaint condition were identified.
 
Event Description
It was reported that during a cholecystectomy, the new ligamax stuck after few firing.Surgery was delayed five minutes, but was successfully completed with a new ligamax.No fragments were generated and there were no patient consequences.No other medical intervention was required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX-5MM ENDO CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kara ditty-bovard
475 calle c
guaynabo 00969
*   00969
6107428552
MDR Report Key9753001
MDR Text Key194450598
Report Number3005075853-2020-01230
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036001843
UDI-Public10705036001843
Combination Product (y/n)N
PMA/PMN Number
K050344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEL5ML
Device Catalogue NumberEL5ML
Device Lot NumberT94U28
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/05/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/29/2020
Initial Date FDA Received02/26/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/26/2019
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-